KevJB
Well-known member
Below are responses from Hayward to my questions concerning differences in salt readings displayed on the Omnilogic display vs. the Taylor K-1766 test kit:
The cell has 2 polarities. It reverses polarity based on the output percentage. The calibration procedure manually does this. When the polarity is reversed, the average salt resets. We don't assume either reading is correct. That's why you need to have it independently verified. Yes, we allow a variance of 500 ppm with our salt chlorinators.
The salt level is calculated by an algorithm in the main board that uses the voltage, amperage and water temperature. The unit may need to be calibrated. Go to the main MSP panel. Press on the chlorinator icon. Press on the magnifying glass icon. Press the +/- icon to calibrate the salt level.
The salt cell you have is an extended life cell. To make these cells extended life the cell plates are over coated with ruthenium, and this will make it read a higher salt level then you have in the pool. As the rethenium wears off the plates the salt levels will come down closer to what you have in the pool.
When using a variable speed pump it is recommended that the cell be plumbed in vertical position. By doing this the cell will fill even when the pump is running in a lower speed. If the cell is plumbed horizontally, and the pump is running in low speed, the cell will not fill completely with water, and cause it to read a false low reading.
Drain a few units from the pool and replenish with fresh water. The coating material is not visibly evident like debris would be. You want to have enough flow to completely cover the internal cell plates. Attempting to operate the unit on a low RPM setting will cause the plates not to be submerged and will give you erratic salt readings.
2800 ppm - that should cover all the cell plates if there's not a blockage to the water flow (like a clogged filter).
The variance is more than 500 ppm. Is the cell new? Some cells leave the factory with an extra coating on the internal cell plates. This causes it to read higher upon installation (usually around 1000 ppm higher). You can remove a few inches of water and replenish with fresh to "trick" the system. As the extra coat wears away, you should start to see the levels normalize. The extra coating will actually extend the life of the cell.
Hayward Case Management Specialist.
Oh the horror!
The cell has 2 polarities. It reverses polarity based on the output percentage. The calibration procedure manually does this. When the polarity is reversed, the average salt resets. We don't assume either reading is correct. That's why you need to have it independently verified. Yes, we allow a variance of 500 ppm with our salt chlorinators.
The salt level is calculated by an algorithm in the main board that uses the voltage, amperage and water temperature. The unit may need to be calibrated. Go to the main MSP panel. Press on the chlorinator icon. Press on the magnifying glass icon. Press the +/- icon to calibrate the salt level.
The salt cell you have is an extended life cell. To make these cells extended life the cell plates are over coated with ruthenium, and this will make it read a higher salt level then you have in the pool. As the rethenium wears off the plates the salt levels will come down closer to what you have in the pool.
When using a variable speed pump it is recommended that the cell be plumbed in vertical position. By doing this the cell will fill even when the pump is running in a lower speed. If the cell is plumbed horizontally, and the pump is running in low speed, the cell will not fill completely with water, and cause it to read a false low reading.
Drain a few units from the pool and replenish with fresh water. The coating material is not visibly evident like debris would be. You want to have enough flow to completely cover the internal cell plates. Attempting to operate the unit on a low RPM setting will cause the plates not to be submerged and will give you erratic salt readings.
2800 ppm - that should cover all the cell plates if there's not a blockage to the water flow (like a clogged filter).
The variance is more than 500 ppm. Is the cell new? Some cells leave the factory with an extra coating on the internal cell plates. This causes it to read higher upon installation (usually around 1000 ppm higher). You can remove a few inches of water and replenish with fresh to "trick" the system. As the extra coat wears away, you should start to see the levels normalize. The extra coating will actually extend the life of the cell.
Hayward Case Management Specialist.
Oh the horror!